Vacuum UV (VUV) Test Facility

The system consists of a high-vacuum chamber, multi-ported, equipped with deuterium and xenon sources for accelerated ageing of coatings under vacuum.

The specimen to be irradiated is mounted on a temperature-controlled platen, which, in turn is mounted on a XYZ J manipulator. Provision is made for the measurement of emissivity and absorptance without breaking vacuum and at the operating temperature.

The chamber is fitted with a magnetically levitated turbo molecular pump and scroll backing pump meaning the system is oil free.

Samples can be loaded into the chamber from within Class 8 clean tent.

  • Maximum operating temperature of sample: 600 degrees C. Specified by customer

  • Sample size: Four 25x25mm

  • Total number and time of emissivity and absorptance measurements specified by the customer

  • Samples are held under high vacuum
     

Test Data Provided

  • Temperature monitored and logged via three calibrated K-- Type thermocouples for the duration of test

  • Emissivity measured to better than +/-0.01

  • Percentage change in absorptance

  • Mass spectrometer data

 
The chamber can be remotely operated and monitored and automatic, over-temperature shutdown systems are in place to protect samples. The control and pump systems are protected by a UPS which allows a controlled shutdown in the event of power failure and notifies TS-Space Systems staff via several mobile communications channels. Emergency chamber sealing and Argon fill systems can also be engaged.

Type: Service
Features radiation:
Electron
Proton
Gamma
UV
Neutron
Heavy ion
Laser
Specimen length [mm]: 50.0
Specimen width/dia [mm]: 50.0
Specimen height/thk [mm]: 5.0
Data acquisition: Temperature, pressure, RGA, Reflectance
Vacuum limit [hPa] [mbar]: 1.3E-5
Flux: MAX 40SC
Project heritage: Solar Orbiter
Cleanroom classes (ISO 14644-1):
ISO Class ≤5
ISO Class 6
ISO Class 7
ISO Class 8
ISO Class 9
Keywords: Vacuum UV VUV, ts space systems, ts-space