Thermal material testing

ISQ LEE EQUIPMENT
Netzsch LFA 457 Thermal diffusivity measurement, 0.01 mm2/s to 1000 mm2/s, once combine with density and specific heat information thermal conductivity can thus be calculated within range 0.1 W/(m.K) to 2000 W/(m.K) in samples with ∅12,7 mm, 0.1 mm to 6 mm thick, within 50ºC to 1000ºC;

Netzsch TMA 402 testing and analysis equipment for dilatometry measurement of samples under the following conditions: up to 30 mm length with a cross section up to 8mmx8mm with a measurement range ± 2.5 mm and a pre-load range from 0,001 N to 3 N within -150ºC to 1550ºC (furnace temperature);

Netzsch STA 449 Combined System TG-DTA-DSC TG (Termo-Gravimetrics), DTA (Differential Thermal Analysis) or DSC (Differential Scanning Calorimetry) including Cp (specific heat calculation) within -150ºC to 1000ºC (furnace temperature) or from 100ºC to 1650ºC (furnace temperature) in a sample of Ø5,1mm x 1,7mm.

Type: Service
Certification:
ISO/IEC 17025
ISO 9001
AS/EN/JISQ 9100
Keywords: Thermal diffusivity, Dilatometry measurement, Specific Heat, Termo-Gravimetrics, Differential Thermal Analysis, Differential Scanning Calorimetry